SciELO - Scientific Electronic Library Online

 
vol.1 número4A Faculdade de Medicina e Cirurgia do Pará: da fundação à federalização 1919-1950Prevalência de enteroparasitismo em crianças de comunidades ribeirinhas do Município de Coari, no médio Solimões, Amazonas, Brasil índice de autoresíndice de assuntospesquisa de artigos
Home Pagelista alfabética de periódicos  

Serviços Personalizados

Journal

Artigo

Indicadores

  • Não possue artigos citadosCitado por SciELO

Links relacionados

  • Não possue artigos similaresSimilares em SciELO

Compartilhar


Revista Pan-Amazônica de Saúde

versão impressa ISSN 2176-6215versão On-line ISSN 2176-6223

Rev Pan-Amaz Saude v.1 n.4 Ananindeua dez. 2010

http://dx.doi.org/10.5123/S2176-62232010000400003 

ORIGINAL ARTICLE

 

Comparison of two automated chemiluminescence tests for the detection of antibodies against the hepatitis C virus

 

 

Felicidade Mota Pereira; Leonardo Assis Bertollo; Maria Alice Sant'Anna Zarife

Laboratório Central de Saúde Pública do Estado da Bahia, Salvador, Bahia, Brasil

Endereço para correspondência
Correspondence
Dirección para correspondencia

 

 


ABSTRACT

INTRODUCTION: A correct diagnosis of hepatitis C virus (HCV) infection is important because treatment is highly expensive and has severe side effects. The assays for the detection of antibodies against HCV (anti-HCV) have undergone several changes, and third-generation tests have been widely used due to their increased sensitivity and specificity.
OBJECTIVE: In this study, we aimed to compare two commercially available, automated third-generation chemiluminescence tests for the detection of anti-HCV antibodies.
METHODS: We analyzed 67 samples from the serum databank from the Laboratório Central de Saúde Pública do Estado da Bahia (LACEN-BA). The reagents examined were the automated ARCHITECT® anti-HCV assay (Abbott Diagnostics, Wiesbaden, Germany) and Elecsys® anti-HCV assay (Roche Diagnostics, Basel, Switzerland) tests and the confirmatory Recombinant Immunoblot Assay (RIBA) test (Chiron RIBA VHC 3.0 SIA, Chiron Corporation, Emeryville, CA, USA).
RESULTS: The Architect anti-HCV assay indicated 18 positive samples, whereas 39 were negative and ten were indeterminate. The Elecsys anti-HCV assay indicated 47 negative and 20 positive samples and no indeterminate samples. Of the ten indeterminate samples indicated by the Architect anti-HCV assay, five were negative by the RIBA test and five were indeterminate. All indeterminate samples indicated by the RIBA test were reactive against the c33 protein.
CONCLUSION: The agreement between the two tests (62 samples) was 91.9%. The Elecsys anti-HCV assay appears to be less sensitive than the Architect anti-HCV, particularly for the detection of the c33 protein shown by the RIBA test. In addition, the Architect anti-HCV assay indicated more indeterminate results out of the negative samples confirmed by the RIBA test, which suggests a lower specificity than the Elecsys anti-HCV assay.

Keywords: Hepatitis C; Hepatitis C Antibodies; Laboratory Test.


 

 

INTRODUCTION

The diagnosis and monitoring of hepatitis C virus (HCV) infection are based on two types of tests: a serological test that detects that detects HCV antigen-specific antibodies and assays that detect viral RNA or HCV core antigens1. However, anti-HCV antibody detection methods do not distinguish between active infections and resolved infections. In addition, false negative results are common because the window period (seroconversion of anti-HCV) is long and lasts between 45 and 68 days2.

False positive results for anti-HCV antibody tests are a well-known problem. This may occur due to interfering factors such as high gamma globulin levels, nephrotic syndrome, liver diseases, autoimmune diseases, viral or parasitic infections or pregnancy in women3. The Centers for Disease Control and Prevention (CDC) estimates that for immunocompetent individuals, approximately 35% of the anti-HCV ELISA immunoassay (EIA) results are false positives4.

Currently, the detection of anti-HCV antibodies is routinely performed using two primary methodologies: ElAs and chemiluminescence immunoassays (CLIAs)5. ElAs detecting anti-HCV antibodies have undergone several stages of development. The third generation test (EIA III) is the most widely used because of its higher sensitivity and specificity, which is due to the greater reactivity to the NS3 region of the virus and the incorporation of the C100 epitopes and antigens from the NS5 region of the viral genome6. However, automated CLIAs have been replacing EIAs because of their practicality, accuracy and greater than 98% specificity; importantly, the sensitivity of automated CLIAs are similar to that of EIAs6.

The presence of anti-HCV antibodies does not always indicate the existence of a current infection; instead, it may represent a false positive result or be indicative of previous resolved acute hepatitis, which corresponds to an immunological scar6. Thus, despite the high specificity of the third generation tests, confirmatory tests must still be performed on samples with low sample/cut-off (S/CO) ratios to avoid false positive results7.

Currently, CLIAs are routinely used in large testing laboratories. Although they are gradually replacing EIAs, a limited number of studies on the comparison and evaluation of these automated CLIAs have been published8.

The objective of this study was to compare two third-generation, fully automated, commercially available CLIAs for the detection of anti-HCV antibodies that are routinely used in the Laboratorio Central de Saude Publico do Estado da Bahia - LACEN-BA (Public Health Central Laboratory of Bahia State).

 

MATERIALS AND METHODS

We tested 67 samples selected from the LACEN-BA databank. When analyzed previously, 18 samples tested positive for anti-HCV antibodies, 39 tested negative and 10 exhibited indeterminate results. The LACEN-BA routinely uses the Architect Anti-HCV Assay Kit (Abbott Diagnostics, Wiesbaden, Germany). In this study, this kit was used as the reference and was compared to the Elecsys Anti-HCV Assay Kit (Roche Diagnostics, Basel IA, Switzerland). A confirmatory   recombinant  immunoblot  assay   (RIBA) (Chiron RIBA HCV 3.0 SIA; Chiron Corporation, Emeryville, CA, USA) was performed on samples with indeterminate results.

The manufacturers' instructions were followed for all tests performed. The Architect anti-HCV assay is a CLIA performed on fully automated equipment (Abbott Architect i4000). Although also performed using completely automated equipment (Roche Modular E 170), the Elecsys anti-HCV assay uses electrochemiluminescence (ECL) technology.

RIBAs utilize strips impregnated with synthetic and recombinant peptides (c100, c33, c22 and NS5) that are applied on separate lines in a solid phase1.

The characteristics of the two automated tests are summarized in table 1. In both methodologies, the emitted signal light is directly proportional to the antibody titer present in that sample.

 

 

The results of both tests are based on a cut-off (CO) value of 1.0. Architect anti-HCV assay results are expressed by the ratio of the sample's optical density to the cut-off measure (S/CO), while Elecsys anti-HCV assay results are presented as index values. The LACEN-BA considers sample results to be anti-HCV antibody positive when they are four times higher than the CO value (> 4.0), negative when they are < 0.9, and indeterminate when they are between 0.9 and 4.0 for both tests. The CO value was established because of its higher concordance (95%) with the RIBA test positive results obtained using the Chiron RIBA test.

 

RESULTS

According to the Architect anti-HCV assay, 18 samples were positive, 39 were negative and 10 were indeterminate; in contrast, 20 samples were positive and 47 were negative according to the Elecsys anti-HCV assay. The Elecsys anti-HCV assay did not report any indeterminate results. These data are summarized in table 2.

 

 

The 10 samples with indeterminate results from the Architect anti-HCV assay had S/CO ratios ranging from 1.12 to 3.86. Of these samples, five were negative and five were indeterminate according to the Chiron RIBA. When analyzed using the Elecsys anti-HCV assay, eight samples were negative and two were positive. Thus, of the 67 samples analyzed, 44 were negative, 18 were positive and 5 exhibited indeterminate results according to the Chiron RIBA. All samples with indeterminate Chiron RIBA results exhibited reactivity to the c33 protein of HCV. The laboratory results of the samples classified as indeterminate by the Architect anti-HCV assay, as well as the results of the RIBA test, are summarized in table 3.

 

 

DISCUSSION

The number of samples evaluated in this study was relatively small compared to other studies that have evaluated the performance of commercially available anti-HCV antibody detection tests1,5,8.

Of the 67 samples analyzed, 10 (14.9%) exhibited discordant results between the two tests evaluated in this study. For the Architect anti-HCV assay, these samples had S/CO ratios of less than 4.0 and were considered indeterminate. However, according to the Elecsys anti-HCV assay, two of these samples were positive and eight were negative. The use of the Chiron RIBA resolved the discordant in 50% of these samples, as five out of 10 tested negative by immunoblot analysis. The samples that exhibited indeterminate results from the Chiron RIBA tested positive only for the c33 protein.

According to CDC recommendations, anti-HCV antibody titer determination requires the use of a screening test with high sensitivity; additionally, reactions with low positivity should be confirmed with a recombinant immunoblot test or a polymerase chain reaction (PCR) test to detect viral RNA8.

Although RIBAs can be used as supplementary confirmatory tests for the detection of anti-HCV antibodies, their disadvantages include a high rate of indeterminate results5.

Berger et al1 also evaluated two serological tests and obtained different results, with a 1.9% discordant between tests, and a resolvability of 63.2% for samples using an immunoblotting test with a reactivity of up to two proteins.

The RIBA indeterminate samples were not tested for viral RNA because of the insufficient remaining material necessary for analysis. However, the presence of anti-HCV antibodies does not necessarily indicate a current infection. Low antibody titers can occur during seroconversion and also when antibody titers decrease following infection resolution.

Patients who have a low risk for hepatitis C infection that have indeterminate RIBA results and a low S/CO ratio typically exhibit undetectable levels of virus when analyzed using PCR7.

When definitely positive and negative samples (62 samples) were compared, the two tests evaluated had a concordance of 91.9%; this result is lower than the concordance reported by Kim et al8. In their study, concordance among four anti-HCV antibody tests, including the Architect anti-HCV assay and the Elecsys anti-HCV assay, ranged from 94.5% to 98.1%.

Following the initial commercial release of the Architect anti-HCV assay, issues with false negative and false positive results were reported. The diluent reagent was subsequently modified, and new studies have shown improved sensitivity and specificity1.

Echevarria et al9 demonstrated that the new Architect anti-HCV assay effectively detects low levels of anti-HCV antibodies, making it an effective screening test in blood banks.

With regards to the advantages and disadvantages of each test, the Elecsys anti-HCV assay exhibits several disadvantages compared to the Architect anti-HCV assay. The test reagents in the Elecsys anti-HCV Assay Kit are not ready for use, creating potential issues regarding contamination   and   pre-analytical   error.   Additionally, calibration is required for each kit, not each lot; thus, as four tests and two controls are used for each calibration, the number of useable tests decreases from 100 to 94 tests per kit. In addition, the period of stability of the reagents in the equipment is three days, and the volume required for each sample is greater compared to the Architect anti-HCV assay.

 

CONCLUSIONS

According to the results obtained in this study, the Elecsys anti-HCV assay was less sensitive compared to the Architect anti-HCV assay regarding the detection of the c33 protein. On the other hand, the Architect anti-HCV assay had a greater number of indeterminate results in negative samples as confirmed using the RIBA test, suggesting lower test specificity compared to the Elecsys anti-HCV assay.

To avoid false positive results, all samples with low S/CO ratios should be confirmed with additional and/or confirmatory tests, such as immunoblotting or PCR tests.

Additional studies are needed to better assess assays that utilize chemiluminescence methodologies. Besides this, these studies should include a larger number of samples.

 

REFERENCES

1 Berger A, Rabenau H, Allwinn R, Doerr HW. Evaluation of the new Architect anti-VHC screening test under routine laboratory conditions. J Clin Virol. 2008 Oct;43(2):158-61. DOI:10.1016/J.JCV.2008.05.009           [ Links ]

2 Morota K, Fujinami R, Kinukawa H, Machida T, Ohno K, Saegusa H, et al. A new sensitive and automated chemiluminescent microparticle immunoassay for quantitative determination of hepatitis C virus core antigen. J Virol Methods. 2009 Apr;157(1):8-14. DOI:10.1016/J.JVIROMET.2008.12.009            [ Links ]

3 Zachary P, Ullmann M, Djeddi S, Meyer N, Wendling MJ, Schvoerer E, et al. Evaluation of three comercially available hepatitis C virus antibody detection assays under the conditions of a clinical virology laboratory. J Clin Virol. 2005 Nov;34(3):207-10. DOI:10.1016/J.JCV.2005.06.005            [ Links ]

4 Watterson JM, Stallcup P, Escamilla D, Chernay P, Reyes A, Trevino SC. Evaluation of the ortho-clinical diagnostics vitros ECi anti-HCVT test: comparison with three other methods. J Clin Lab Anal. 2007;21(3):162-6.            [ Links ]

5 Kesli R, Ozdemir M, Kurtoglu MG, Baykan M, Baysal B. Evaluation and comparison of three different anti-hepatitis C virus antibody tests based on chemiluminescence and enzyme-linked immunosorbent assay methods used in the diagnosis of hepatitis C infections in Turkey. J Int Med Res. 2009 Sep-Oct;37(5):1420-9.            [ Links ]

6 Parise E. Diagnóstico sorológico das hepatites crônicas virais. Roche Rev. 2009;8:2-6.

7 Vermeersch P, Van Ranst M, Lagrou K. Validation of a strategy for HCV antibody testing with two enzyme immunoassays in a routine clinical laboratory. J Clin Virol. 2008 Aug;42(4):394-8. DOI:10.1016/J.JCV.2008.02.015             [ Links ]

8 Kim S, Kim JH, Yoon S, Park YH, Kim HS. Clinical performance evaluation of four automated chemiluminescence immunoassays for hepatitis C virus antibody detection. J Clin Microbiol. 2008 Dec;46(12):3919-23. DOI:10.1128/JCM.01603-08            [ Links ]

9 Echevarria JM, Avellón A, Jonas G, Hausmann M, Vockel A, Kapprell HP. Sensitivity of a modified version of the ARCHITECT® anti-HCV test in detecting samples with immunoblot-confirmed, low-level antibody to hepatitis C virus. J Clin Virol. 2006 Apr;35(4):368-72. DOI:10.1016/J.JCV.2005.11.006             [ Links ]

 

 

Correspondência / Correspondence / Correspondencia:
Felicidade Mota Pereira
Rua Waldemar Falcão, 123. Bairro: Candeal.
CEP: 40296-710 Salvador-Bahia-Brasil
Tel: +55 (71) 3276-1721
E-mail: felizmp@yahoo.com.br

Recebido em / Received / Recibido en: 9/8/2010
Aceito em / Accepted / Aceito en: 20/12/2010